Feeding
Flow rate range:
- 1 – 100 mL/h (1 mL/h increments)
- 100 – 600 mL/h (5 mL/h increments)
Flow rate accuracy:
Feeding mode:
Priming:
- Automatic, semi-automatic and manual priming are available
Counter:
- Cumulative feeding volume counter from 0.001 L to 99.999 L
Data event log:
- Feeding history: 250 events
- Alarm history: 250 events
Night mode:
- The night mode decreases the brightness of the screen and the power LED
Keypad lock:
- Disabling keypad to prevent unintentional key press
Setting lock:
- Possibility to prevent patient from changing settings, access code protected
Other settings:
- Target volume activation/deactivation, sound level adjustment, key beep activation/deactivation, contrast & brightness settings, time between alarm sounds setting, time between target volume almost reached message and target volume reached setting
Intented use:
- For enteral feeding use only
Information / Alarms
Pump status:
- Administration in progress: Droplet animation
- Administration stopped: Stop symbol displayed
- All alarms represented by light indicators, pictograms and sounds
Information:
- Target volume almost reached, battery almost discharged, start reminder, technical information
Alarms:
- Target volume reached, door opened, wrong set installation, downstream occlusion, upstream occlusion, empty bag / air in line, empty battery, technical error
Occlusion detection time:
- At 50 mL/h: < 6 min with medical water
Empty bag / air in line detection time:
- At 50 mL/h: < 5 min with medical water
Technical Specifications
Pumping mechanism: Linear peristaltic pumping system
Display: Monochrome LCD 40 x 59 mm (160 x 240 pixels)
Clamp: Versatile clamp to fix on a rail (10 - 35 mm) or a pole (Ø 8 - 40 mm)
Dimensions (H/W/D) / Weight:
- Pump: 138 x 128 x 48 mm / 610 g
- Pump Holder: 185 x 110 x 85 mm / 350 g
Battery: NiMH (Nickel-Metal Hydride), battery life 24 h ± 5 % at 125 mL/h, battery level indication on the display
- Battery charging time: 6 h
Compliance standard:
- IEC 60 601-1, IEC 60 601-1-2 and IEC 60601-2-24
- CE 0123 marking in compliance with the EEC 93/42 European Medical Device Directive
Waterproofness:
- Pump: IP35
- Pump Holder: IP32
- AC Adaptor: IP41
Electrical compliance:
- Protection against leakage current: Defibrillation-proof type CF
- Protection against electric shocks: class II
Power supply: 100 – 240 VAC / 50 - 60 Hz
Power consumption: Maximum 7 W in standard operating conditions
Noise: Around 25.8 dBA at 1 m (background noise 18.3 dBA)
Housing material: ABS
Amika Administration Set
Free-flow protection: All sets are designed with the Amika clamp preventing the risk of free-flow when the door is opened and when the set is disengaged
Set length: Around 2.50 m, adjustable anti-free-flow Amika clamp positioning
Materials: All Amika lines administration sets are DEHP-free & latex-free